首页 >>> 产品目录 >>> 通用仪器产品 >>> 模拟示波器
|
|
产品[示波器(100MHz)]资料
|
如果您对该产品感兴趣的话,可以
产品名称:
示波器(100MHz)
产品型号:
OS-5100
产品展商:
国营建华仪器厂
简单介绍
100MHz, 2Ch, Dual trace, Delay sweep function
DC to 100MHz bandwidth, 2 channels, Dual trace
6" rectangular CRT with internal graticule
Wide dynamic range even at high frequency
ALT triggering function(VERT mode)
Auto focusing according to the change of intensity
Drift compensation circuit employed in vertical amplifier for low drift
Jitterless and high trigger sensitivity
TV sync. Separator circuit for stable TV signal observation
X-Y operation
Signal delay line installed
|
|
示波器(100MHz)的详细介绍 |
CRT
|
- Configuration and useful screen : 6 inch rectangular screen with internal graticule, 8X10div. (1div = 1cm) marking for measurement of rise time, 2mm subdivisions along the central axis. - Acceleration potential : +10.5kv approx.(ref. Cathode) - Phosphor : P31 - Focusing : provided - Trace rotation : provided - Intensity control : provided
|
|
|
|
|
Z-AXIS INPUT(INTENSITY MODULATION)
|
- Input signal : positive going signal decreases +5Vp-p or more signal cases intensity noticeable modulation at normal intensity setting - Bandwidth : DC to 3.5MHz(-3dB) - Coupling : DC - Input impedance : 17~22kз typical - Maximum input voltage : 20V(DC + peak AC)
|
|
|
|
|
VERTICAL DEFLECTION
|
- DC coupled(-3dB) : DC to 100MHz normal / DC to 20MHz magnified - AC coupled(-3dB) : 10Hz to 100MHz normal / 10Hz to 20MHz magnified - Modes : CH1, CH2, ADD, CHOP, ALT - Deflection factor (normal): 2mV/div. to 5V/div. in 11 calibrated steps of 1-2-5 sequence. Continuously variable between steps at least 1:2.5 X 5. - Accuracy : normal ±3% - Input impedance : approx. 1Mз in parallel with 25pF - Maximum input voltage : direct 250V (DC + peak AC) - Input coupling : AC, GND, DC - Rise time : 3.5ns or less (17.5ns or less : 2mV/div.) - CH1 output : 20mV/div. into 50з / 50Hz to 30MHz(-3dB) - Polarity inversion : CH2 only - Signal delay : delay cable installed
|
|
|
|
|
HORIZONTAL DEFLECTION
|
- Sweep mode : A, A int B, B, B triggered, X-Y - Time base A : 0.1Я/div. to 0.5s/div. in 21 calibrated steps of 1-2-5 sequence uncalibrated continuous control between steps at least 1:2.5 - Hold-off time on time base A : variable with the hold-off control - Time base B : 0.1Я/div. to 10Я/div. in 7 calibrated steps of 1-2-5 sequence - Delay sweep on time base B: 1div. or less / 10 div. or more - Sweep magnification : 10 times (max. sweep rate : 10ns/div.) - Accuracy : ±3%, ±5%(1:10 probe)
|
|
|
|
|
TRIGGER SYSTEM
|
- Accuracy : - Modes : AUTO, NORM, TV-V, TV-H - Source : INT, LINE, EXT - Coupling : AC - Slope : + or - - Sensitivity and frequency in AUTO, NORM 30Hz -10MHz : 0.48div. in INT / 200mVp-p in EXT - Sensitivity and frequency in AUTO, NORM 10MHz -100MHz : 1.5div. in INT / 600mVp-p in EXT - Sensitivity and frequency in ALT trigger 30Hz -10MHz : 1.5div. in INT - Sensitivity and frequency in ALT trigger 10MHz -10MHz : 1.5div. in INT - Maximum input voltage : 250V (DC + AC peak)
|
|
|
|
X-Y OPERATION
|
- Sensitivity : same as vertical deflection for both X-axis(CH1) except accuracy ±5% in X-axis - Frequency and bandwidth : DC to 2MHz (DC coupling) / 10Hz to 2MHz (AC coupling) - X-Y phase difference : 3°or less as DC to 50kHz
|
|
|
|
CALIBRATOR
|
- Probe adjustment : approx. 1kHz frequency, 0.5V(±3%) square wave with 50% duty ratio
|
|
|
|
POWER SUPPLY
|
- Line voltage range : 100V(90-110V) with 250V 2A fuse, 120V(108-132V) with 250V 2A fuse, 220V(198-242V) with 250V 1A fuse or 230V(216-250V) with 250V 1A fuse - Line frequency : 50/60Hz - Power consumption : approx. 60WOTHERSz
|
|
|
|
PHYSICAL
CHARACTERISTICS
|
- Weight : 9.6kg - Dimension : 320mm(W) X 140mm(H) X 430mm(L)
|
|
|
|
|
OTHERS
|
- Accessories including : Operation manual, Spare fuse, Power cord - Optional accessories (sold separately) : Probe
|
|
|
|
|